Trade-off between the radiation parameters and image quality using iterative reconstruction techniques in head
Elīza Lāce1, Reza Mohammadian1, Ainārs Āboltiņš2
1Department of Radiology, Riga Stradin's University, Riga, Latvia.
Background:
Iterative reconstruction techniques (IRTs) are commonly used in computed tomography (CT) and help to reduce image noise.
Purpose:
To determine the minimum radiation dose while preserving image quality in head CT using IRTs.
Material And Methods:
The anthropomorphic phantom was used to scan nine head CT image series with varied radiation parameters. CT dose parameters, including volume CT dose index (CTDIvol [in mGy]) and dose length product (DLP [in mGy/cm]), were recorded for each scan series. Different noise levels (iDoseL1-6) were used in IRT reconstructions for soft and bone tissues. In total, 15 measurements were taken from five regions of interest (ROI) with an area of 10 mm2. The signal-to-noise ratio (SNR) and noise values obtained at different ROIs were compared among various reconstruction methods with repeated measures of statistical analysis.
Results:
In the head CT scan, applying IRT iDoseL5 had the lowest noise and highest SNR for soft tissue (P < 0.05), and increased iDose can decrease CT dose by 54.6% without compromising image quality. While for bone tissue reconstruction, no clear association was found between the level of iDose and noise. However, when CTDIvol is >20 mGy, iDoseL4 is slightly superior to other reconstruction methods (P < 0.065).
Conclusion:
Using IRTs in head CTs reduces radiation dose while maintaining image quality. IDoseL5 provided optimal balance for soft tissue.
More Related Videos
05:32Retrospective Cardiac Gating with A Prototype Small-Animal X-ray Computed Tomograph
Published on: February 21, 2025
06:53Management of Respiratory Motion Artefacts in 18F-fluorodeoxyglucose Positron Emission Tomography using an Amplitude-Based Optimal Respiratory Gating Algorithm
Published on: July 23, 2020
Related Concept Videos
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...