Problem Solving MRI to Reduce False-Positive Biopsy Related to Breast US: Conductivity vs. DWI vs. Abbreviated
Jun-Hyeong Kim1, Soo-Yeon Kim2,3,4, Chuanjiang Cui1
1Department of Electrical and Electronic Engineering, Yonsei University, Seoul, Republic of Korea.
Background:
While breast ultrasound (US) is a useful tool for diagnosing breast masses, it can entail false-positive biopsy results because of some overlapping features between benign and malignant breast masses and subjective interpretation.
Purpose:
To evaluate the performance of conductivity imaging for reducing false-positive biopsy results related to breast US, as compared to diffusion-weighted imaging (DWI) and abbreviated MRI consisting of one pre- and one post-contrast T1-weighted imaging.
Study Type:
Prospective.
Subjects:
Seventy-nine women (median age, 44 years) with 86 Breast Imaging Reporting and Data System (BI-RADS) category 4 masses as detected by breast US.
Field Strength/Sequence:
3-T, T2-weighted turbo spin echo sequence, DWI, and abbreviated contrast-enhanced MRI (T1-weighted gradient echo sequence).
Assessment:
US-guided biopsy (reference standard) was obtained on the same day as MRI. The maximum and mean conductivity parameters from whole and single regions of interest (ROIs) were measured. Apparent diffusion coefficient (ADC) values were obtained from an area with the lowest signal within a lesion on the ADC map. The performance of conductivity, ADC, and abbreviated MRI for reducing false-positive biopsies was evaluated using the following criteria: lowest conductivity and highest ADC values among malignant breast lesions and BI-RADS categories 2 or 3 on abbreviated MRI.
Statistical Tests:
One conductivity parameter with the maximum area under the curve (AUC) from receiver operating characteristics was selected. A P-value <0.05 was considered statistically significant.
Results:
US-guided biopsy revealed 65 benign lesions and 21 malignant lesions. The mean conductivity parameter of the single ROI method was selected (AUC = 0.74). Considering conductivity (≤0.10 S/m), ADC (≥1.60 × 10-3 mm2 /sec), and BI-RADS categories 2 or 3 reduced false-positive biopsies by 23% (15 of 65), 38% (25 of 65), and 43% (28 of 65), respectively, without missing malignant lesions.
Data Conclusion:
Conductivity imaging may show lower performance than DWI and abbreviated MRI in reducing unnecessary biopsies.
Evidence Level:
2 TECHNICAL EFFICACY: Stage 2.
More Related Videos
06:08A Cognitive Fusion-guided Prostate Biopsy Using Multiparametric Magnetic Resonance Imaging and Transrectal Ultrasound
Published on: March 21, 2025
15:48Tracking the Mammary Architectural Features and Detecting Breast Cancer with Magnetic Resonance Diffusion Tensor Imaging
Published on: December 15, 2014
Related Concept Videos
Magnetic Resonance Imaging
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Imaging Studies IV: Magnetic Resonance Imaging
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...