A Pneumatic Low-Pass Filter for High-Fidelity Cuff-Based Pulse Waveform Acquisition
Alessio Tamborini1, Morteza Gharib2
1California Institute of Technology, 1200 E California BLVD MC 105-50, Pasadena, CA, 91125, USA. atambori@caltech.edu.
Abstract:
Cuff-based pulse waveform acquisition (CBPWA) devices are reliable solutions for non-invasive cardiovascular diagnostics. However, poor signal resolution has limited clinical applications. This study aims to demonstrate the improved signal quality of CBPWA devices by implementing passive pneumatic low-pass filters (pLPF). Conventionally, pressure sensor output resolution is a percentage of the operating range. Therefore, measurement of small pressure changes in a large range must sacrifice signal resolution to accommodate for the large mean pressures. We design a pLPF to obtain the running mean pressure and combine it with a high-resolution differential pressure sensor for isolating the signal's pulsatile component. Thirty-one volunteers participated in a device proof-of-concept study at Caltech. Volunteers were measured at rest in the supine position on the left arm. The filtering behavior is mathematically modeled and experimentally verified, showing good agreement between measured and predicted cutoff frequencies. In the human study, the device successfully captured high-fidelity pulse waveform measurements for all volunteers: a blood pressure (BP) reading was followed by inflate-and-hold acquisition in diastolic BP (DBP), mean arterial pressure (MAP), and supra systolic BP (sSBP). The study demonstrated the reliability and high signal resolution of pLPF for CBPWA. Considering the widespread use of the brachial cuff, a system for high-resolution CBPWA motivates the clinical implementation of non-invasive pulse waveform analysis (PWA).
Related Concept Videos
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Passive Filters
Low-Pass Filters
Low-pass filters are designed to transmit signals with frequencies lower than the cutoff frequency, ωc, and attenuate those above it. The cutoff...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Pulse
The pulse serves as a clinical...