Thermal Profiles for Focal Pulsed Electric Field Ablation
Atul Verma1, Princeton Zhong2, Quim Castellvi2
1Division of Cardiology, McGill University Health Centre, McGill University, Montreal, Quebec, Canada.
Background:
Pulsed electrical field (PEF) ablation may cause tissue heating. These changes are reportedly small, but each PEF system and waveform will have a different behavior, and data are lacking.
Objectives:
This study sought to compare the temperature profile of focal point, monopolar biphasic PEF ablation versus radiofrequency (RF).
Methods:
Ablation lesions were performed on perfused thigh muscle of swine. PEF lesions were performed with 3 compatible ablation catheters at the highest (25 amp) energy, and 1 catheter (Tacticath SE) was also used at the 22- and 19-amp levels. Temperature changes in the tissue were measured using fluoroptic temperature probes inserted at the muscle surface, as well as 3 mm and 7 mm below the surface. Temperatures were recorded continuously at baseline, during delivery, and after ablation. Muscle temperatures were compared with those of RF lesions performed with 1 catheter (Tacticath SE) at 30 W for 30 seconds.
Results:
PEF ablation with 3energy settings produced small temperature changes. Maximum average temperature rise for PEF for the maximum (25-amp) energy setting (32 lesions) was 7.6 °C, 2.8 °C, and 0.9 °C at the surface, 3-mm depth, and 7-mm depth, respectively. The temperature rise was dose dependent, with lower energy settings yielding less temperature rise. RF ablations (10 lesions) produced temperature increases of 16.6 °C, 39.8 °C, and 9.5 °C at the surface, 3-mm depth, and 7-mm depth, respectively.
Conclusions:
PEF caused detectable temperature changes in muscle tissue, which never exceeded 2.8 °C at the 3-mm depth versus baseline. By contrast, RF produced substantial temperature rises. These data support that focal monopolar biphasic energy delivered by this PEF technology retains a favorable thermal safety profile.