Sex differences in leadless pacemaker implantation: A propensity-matched analysis from the i-LEAPER registry
Gianfranco Mitacchione1, Marco Schiavone2, Alessio Gasperetti3
1Department of Cardiology, Luigi Sacco University Hospital, Milan, Italy.
Background:
The impact of sex in clinical and procedural outcomes in leadless pacemaker (LPM) patients has not yet been investigated.
Objective:
The purpose of this study was to investigate sex-related differences in patients undergoing LPM implantation.
Methods:
Consecutive patients enrolled in the i-LEAPER registry were analyzed. Comparisons between sexes were performed within the overall cohort using an adjusted analysis with 1:1 propensity matching for age and comorbidities. The primary outcome was the comparison of major complication rates. Sex-related differences regarding electrical performance and all-cause mortality during follow-up were deemed secondary outcomes.
Results:
In the overall population (n = 1179 patients; median age 80 years), 64.3% were men. After propensity matching, 738 patients with no significant baseline differences among groups were identified. During median follow-up of 25 [interquartile range 24-39] months, female sex was not associated with LPM-related major complications (hazard ratio [HR] 2.03; 95% confidence interval [CI] 0.70-5.84; P = .190) or all-cause mortality (HR 0.98; 95% CI 0.40-2.42; P = .960). LPM electrical performance results were comparable between groups, except for a higher pacing impedance in women at implant and during follow-up (24 months: 670 [550-800] Ω vs 616 [530-770] Ω; P = .014) that remained within normal limits.
Conclusion:
In a real-world setting, we found differences in sex-related referral patterns for LPM implantation with an underrepresentation of women, although major complication rate and LPM performance were comparable between sexes. Female patients showed higher impedance values, which had no impact on overall device performance. Electrical parameters remained within normal limits in both groups during the entire follow-up.