Artefactually low creatinine by Beckman Coulter enzymatic method due to immunoglobulin M paraprotein interference
Bobby Li1,2, Simon Thompson1,2, Kareen Igot1,2
1Te Whatu Ora - Health New Zealand Waitaha Canterbury, Christchurch, New Zealand.
Abstract:
An 81-year-old man was admitted to hospital with symptomatic coronavirus disease (COVID-19) infection. He had a background of progressive chronic inflammatory demyelinating polyneuropathy associated with Waldenstrom's macroglobulinaemia. His plasma creatinine on four separate samples was inconceivably low (all ≤13 μmol/L), as measured by a Beckman Coulter enzymatic assay) after being 72 μmol/L 3 months earlier. On further investigation, his serum immunoglobulin M (IgM) was 15.4 g/L and his plasma creatinine measured by Roche enzymatic and Roche Jaffe methods was 62 μmol/L and 64 μmol/L, respectively. This was consistent with results post dilution studies and polyethylene glycol (PEG) precipitation on the Beckman Coulter assay. There was no evidence of similar interference when reviewing creatinine results from 10 other patients with IgM paraproteinaemia who had been tested in our laboratory. Clinicians and laboratorians are reminded that enzymatic creatinine is not free from interferences. IgM paraprotein negative interference of enzymatic creatinine is rare and specific to a patient's IgM and assay combination, but should be considered in patients with an unexplained low enzymatic creatinine result. Useful investigations to identify an interference include dilution studies, PEG precipitation and measuring creatinine on an alternative method such as Jaffe, mass spectrometry or an enzymatic method from a different platform.
More Related Videos
09:43Analyses of Proteinuria, Renal Infiltration of Leukocytes, and Renal Deposition of Proteins in Lupus-prone MRL/lpr Mice
Published on: June 8, 2022
09:32Procedure and Key Optimization Strategies for an Automated Capillary Electrophoretic-based Immunoassay Method
Published on: September 10, 2017
Related Concept Videos
Serum Studies: Renal Function Tests
Factors Affecting Renal Clearance: Renal Impairment
One condition associated with renal failure is uremia. Uremia is characterized by impaired glomerular filtration and fluid accumulation in the body. This condition hinders the renal clearance of drugs, resulting in drug accumulation and potential...
Enzyme-Linked Immunosorbent Assay
There are many different types of ELISAs, but they all involve an antibody molecule whose constant region binds an enzyme, leaving the variable region free to bind its specific antigen. Enzyme-substrate reaction allows the antigen to be visualized or...
Renal Clearance
Renal clearance refers to the volume of plasma cleared of a specific substance, such as creatinine, per unit of time. To measure clearance, urine samples are collected over a 24-hour period during each bladder voiding, followed by a single blood sample at the...