Proton Therapy Delivery and Its Clinical Application in Select Solid Tumor Malignancies
Published on: February 6, 2019
Synchronization of light flash with the irradiation pulse in proton beam therapy: A case report
Takashi Saito1, Masashi Mizumoto1, Yoshiko Oshiro1,2
1Department of Radiation Oncology, University of Tsukuba, Tsukuba, Japan.
Abstract:
The correlation between sensory light flash and proton beam delivery was evaluated by measuring the timing of pulse beam delivery and light flash sensing using an event recorder in an 83-year-old patient receiving proton beam therapy (PBT) for nasopharyngeal adenoid cystic carcinoma. The treatment dose was 65 Gy (RBE) in 26 fractions with 2 ports, and both beams included the visual pathway (retina, optic nerve, chiasma). Measurements were obtained in 13 of the 26 fractions. The patient sensed a light flash in all 13 fractions and pressed the recorder button for 426 of the 430 pulsed beam deliveries, giving a sensing rate of 99.1%. The median duration of button-pressing of 0.3 s was almost the same as that of the beam pulse of 0.2 s, with a reaction time lag of 0.35 s. These results suggest a consistency between light flash during PBT and the timing of irradiation.