[Sinogram interpolation combined with unsupervised image-to-image translation network for CT metal artifact
1School of Biomedical Engineering, Southern Medical University, Guangzhou 510515, China.
Objective:
To propose a framework that combines sinogram interpolation with unsupervised image-to-image translation (UNIT) network to correct metal artifacts in CT images.
Methods:
The initially corrected CT image and the prior image without artifacts, which were considered as different elements in two different domains, were input into the image transformation network to obtain the corrected image. Verification experiments were carried out to assess the effectiveness of the proposed method using the simulation data, and PSNR and SSIM were calculated for quantitative evaluation of the performance of the method.
Results:
The experiment using the simulation data showed that the proposed method achieved better results for improving image quality as compared with other methods, and the corrected images preserved more details and structures. Compared with ADN algorithm, the proposed algorithm improved the PSNR and SSIM by 2.4449 and 0.0023 when the metal was small, by 5.9942 and 8.8388 for images with large metals, and by 8.8388 and 0.0130 when both small and large metals were present, respectively.
Conclusion:
The proposed method for metal artifact correction can effectively remove metal artifacts, improve image quality, and preserve more details and structures on CT images.
More Related Videos
Related Concept Videos
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System V: CT