In vitro Assessment of Aortic Regurgitation Using Four-Dimensional Flow Magnetic Resonance Imaging
Published on: February 25, 2022
4Dflow-VP-Net: A deep convolutional neural network for noninvasive estimation of relative pressures in stenotic flows
Ruponti Nath1,2, Amirkhosro Kazemi1,2, Sean Callahan1,2
1Medical Imaging Lab, Department of Electrical and Computer Engineering, University of Louisville, Louisville, Kentucky, USA.
Purpose:
To estimate relative transvalvular pressure gradient (TVPG) noninvasively from 4D flow MRI.
Methods:
A novel deep learning-based approach is proposed to estimate pressure gradient across stenosis from four-dimensional flow MRI (4D flow MRI) velocities. A deep neural network 4D flow Velocity-to-Presure Network (4Dflow-VP-Net) was trained to learn the spatiotemporal relationship between velocities and pressure in stenotic vessels. Training data were simulated by computational fluid dynamics (CFD) for different pulsatile flow conditions under an aortic flow waveform. The network was tested to predict pressure from CFD-simulated velocity data, in vitro 4D flow MRI data, and in vivo 4D flow MRI data of patients with both moderate and severe aortic stenosis. TVPG derived from 4Dflow-VP-Net was compared to catheter-based pressure measurements for available flow rates, in vitro and Doppler echocardiography-based pressure measurement, in vivo.
Results:
Relative pressures calculated by 4Dflow-VP-Net and in vitro pressure catheterization revealed strong correlation (r2 = 0.91). Correlations analysis of TVPG from reference CFD and 4Dflow-VP-Net for 450 simulated flow conditions showed strong correlation (r2 = 0.99). TVPG from in vitro MRI had a correlation coefficient of r2 = 0.98 with reference CFD. 4Dflow-VP-Net, applied to 4D flow MRI in 16 patients, showed comparable TVPG measurement with Doppler echocardiography (r2 = 0.85). Bland-Altman analysis of TVPG measurements showed mean bias and limits of agreement of -0.20 ± 2.07 mmHg and 0.19 ± 0.45 mmHg for CFD-simulated velocities and in vitro 4D flow velocities. In patients, overestimation of Doppler echocardiography relative to TVPG from 4Dflow-VP-Net (10.99 ± 6.77 mmHg) was observed.
Conclusion:
The proposed approach can predict relative pressure in both in vitro and in vivo 4D flow MRI of aortic stenotic patients with high fidelity.
More Related Videos
11:00Experimental Investigation of Secondary Flow Structures Downstream of a Model Type IV Stent Failure in a 180° Curved Artery Test Section
Published on: July 19, 2016
09:57Development and Evaluation of 3D-Printed Cardiovascular Phantoms for Interventional Planning and Training
Published on: January 18, 2021
Related Concept Videos
Uniform Depth Channel Flow
Uniform Depth Channel Flow: Problem Solving
Pressure Variation in a Fluid at Rest
When measuring pressure at two different levels within the fluid, the difference in...
Magnetic Resonance Imaging
Measurement of Fluid Pressure
A basic form of manometer is the piezometer, a vertical tube open at the top and filled with the same...