Evaluation of the Partial Re-Sheathing Technique with the Solitaire Stent Retrieval System In Vitro Model and a
Yoshihisa Matsumoto1, Yui Nagata1, Takuro Hashikawa1
1Department of Neurosurgery, St. Mary's Hospital, Kurume, Fukuoka, Japan.
Objective:
Vascular injuries are severe complications associated with endovascular thrombectomy. In the present study, we evaluated the re-sheathing technique with the Solitaire stent retrieval system to overcome these complications.
Methods:
We examined the diameter and resistance to retrieval of the Solitaire FR device (6 × 20 mm) during full and partial deployment in vitro model. We also examined a representative case in which the re-sheathing technique was used.
Results:
We found that the Solitaire device spread elliptically during partial deployment. As the length of the partially deployed device decreased, the maximum diameter also decreased. The distal half of the stent retained 80% of the maximum diameter of the partially deployed Solitaire. The resistance to retrieval was significantly higher during full deployment (mean ± standard deviation; 0.32 ± 0.04 kg) than during half deployment (0.22 ± 0.04 kg) (Mann-Whitney U test; p = 0.006). The re-sheathing technique was used in the representative case due to the high resistance to retrieval, which enabled recanalization without extravasation.
Conclusion:
In cases of high resistance to retrieval, minimal re-sheathing may be useful for capturing the thrombus without increasing the risk of vascular injury.