Investigating Stress-relaxation and Failure Responses in the Trachea
Published on: October 18, 2022
Experimental Study of Coil Compaction: Impact of Pulsatile Stress
Takumi Asai1, Yoshitaka Nagano2, Tomotaka Ohshima3
1Department of Neurosurgery, National Hospital Organization, Nagoya Medical Center, Nagoya, Aichi, Japan.
Objective:
Coil compaction after aneurysm embolization is one of the major issues associated with aneurysm recurrence. On the presumption that pulsatile stress to the aneurysm is responsible for coil compaction, we developed an experimental model in vitro to visualize the mechanical stresses exerted by blood pressure and pulse and their relation to coil compaction.
Methods:
A closed-type non-circulation system was developed by installing a syringe that generated pressure at one end of a tube, along with a spherical aneurysm made of silicone and a pressure sensor in the bifurcated end. We installed a fixed-pressure model under a steady pressure of 300 mmHg while the pressure-fluctuation model simulated the pressure variations using a plunger (in a syringe) by using a motor that applied pulsatile stress in the range of 50 mmHg for a 10-ms cycle. We devised four types of aneurysms with different depths and the same coil length. After coil packing, the aneurysms were observed for 3 days (the observation period in the pressure-fluctuation model corresponded to approximately 300 days in real time). The distance from the datum point to the observable coil loops was determined as the initial position, and the temporal change in the distance from that position was measured.
Results:
In the fixed-pressure model, the average distance of coil movement was very small (less than ±0.1 mm). In the pressure-fluctuation model, the movement of coils was observed to be significant for the two longest depths (0.11 and 0.14 mm). The maximal dynamic change in coil movement was observed on the second day. The range of movement was observed to decrease thereafter.
Conclusion:
Our experimental study enabled the observation of coil movement within a short duration. It examined coil compaction by applying pulsed pressure to the coils at high speeds. Consequently, a shift of the coil loops inside the incompletely occluded aneurysms was detected on applying a pulsed pressure.
More Related Videos
Related Concept Videos
Stresses under Combined Loadings
The process begins by slicing the tube at critical points and analyzing the internal forces and stress components at these sections, focusing on the centroid. Normal stresses, generated by axial forces and bending moments, are either compressive or tensile and vary across the section from...
Stress Concentrations in Circular Shafts
Stress: General Loading Conditions
The shearing force, possessing potential directionality within the plane of the section, is simplified into two component forces running parallel to the x and y axes....
True Stress and True Strain
In contrast, true stress offers a more precise portrayal. It is computed by dividing the...
Residual Stresses
Normal Strain under Axial Loading