Reflection-Boosted Wearable Ring-Type Pulse Oximeters for SpO2 Measurement with High Sensitivity and Low Power
Min Gyu Joo1, Dae Hyeong Lim2,3, Kyu-Kwan Park1
1Department of Electrical and Computer Engineering, Sungkyunkwan University, Suwon 16419, Republic of Korea.
Abstract:
In this study, we demonstrated a Monte Carlo simulation to model a finger structure and to calculate the intensity of photons passing through tissues, in order to determine optimal angular separation between a photodetector (PD) and a light-emitting diode (LED), to detect SpO2. Furthermore, our model was used to suggest a mirror-coated ring-type pulse oximeter to improve the sensitivity by up to 80% and improve power consumption by up to 65% compared to the mirror-uncoated structure. A ring-type pulse oximeter (RPO) is widely used to detect photoplethysmography (PPG) signals for SpO2 measurement during sleep and health-status monitoring. Device sensitivity and the power consumption of an RPO, which are key performance indicators, vary greatly with the geometrical arrangement of PD and LED within the inner surface of an RPO. We propose a reflection-boosted design of an RPO to achieve both high sensitivity and low power consumption, and determine an optimal configuration of a PD and LED by performing a 3D Monte Carlo simulation and confirming its agreement with experimental measurement. In order to confirm the reflection-boosted performance in terms of signal-to-noise ratio, R ratio, and perfusion index (PI), RPOs were fabricated with and without a highly reflective coating, and then used for SpO2 measurement from eight participants. Our simulation allows the numerical calculation of the intensity of photon passing and scattering through finger tissues. The reflection-boosted RPO enables reliable measurement with high sensitivity, resulting in less power consumption for the LED and longer device usage than conventional RPOs without any reflective coating, in order to maintain the same level of SNR and PI. Compared to the non-reflective reference RPO, the reflection-boosted RPO design greatly enhanced both detected light intensity (67% in dc and 322% in ac signals at a wavelength λ1 = 660 nm, and also 81% and 375% at λ2 = 940 nm, respectively) and PI (23.3% at λ1 and 25.5% at λ2). Thus, the reflection-boosted design not only enhanced measurement reliability but also significantly improved power consumption, i.e., by requiring only 36% and 30% power to drive the LED sources with λ1 and λ2, respectively, to produce the device performance of a non-reflective RPO reference. It is expected that our proposed RPO provides long-term monitoring capability with low power consumption and an enhanced PI for SpO2 measurement.
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Special considerations while measuring pulse
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...