Related Experiment Video
Updated: Jul 21, 2025

The Use of a β-lactamase-based Conductimetric Biosensor Assay to Detect Biomolecular Interactions
Published on: February 1, 2018
Bio-Inspired Molecularly Imprinted Polymer Electrochemical Sensor for Cortisol Detection Based on O-Phenylenediamine
Minwoo Kim1, Daeil Park1, Joohyung Park1
1Department of Biomechatronic Engineering, College of Biotechnology and Bioengineering, Sungkyunkwan University, Suwon 16419, Republic of Korea.
Abstract:
This paper presents a comprehensive investigation of the various parameters involved in the fabrication of a molecularly imprinted polymer (MIP) sensor for the detection of cortisol. Parameters such as monomer concentration, electropolymerization cycles, pH, monomer-template ratio, template removal technique, and rebinding time were optimized to establish a more consistent and effective method for the fabrication of MIP sensors. Under the optimized conditions, the MIP sensor demonstrated a proportional decrease in differential pulse voltammetry peak currents with increasing cortisol concentration in the range of 0.1 to 100 nM. The sensor exhibited excellent sensitivity, with a limit of detection of 0.036 nM. Selectivity experiments using a non-imprinted polymer sensor confirmed the specific binding affinity of the MIP sensor for cortisol, distinguishing it from other steroid hormones. This study provides crucial insights into the development of a reliable and sensitive strategy for cortisol detection using O-PD-based MIPs. These findings laid the foundation for further advancements in MIP research.