Redo-TAVI with SAPIEN 3 in SAPIEN XT or SAPIEN 3 - impact of pre- and post-dilatation on final THV expansion
David Meier1,2,3, Uri Landes4,5, Lars Sondergaard6
1Centre for Cardiovascular Innovation, St Paul's Hospital, Vancouver, BC, Canada and Vancouver General Hospital, Vancouver, BC, Canada.
Redo-transcatheter aortic valve implantation (TAVI) with balloon-expandable valves risks underexpansion. Pre- and post-dilatation improved expansion, but only SAPIEN 3 in SAPIEN 3 achieved full expansion with both.
Area of Science:
- Cardiovascular interventions
- Medical device engineering
- Interventional cardiology
Background:
- Redo-transcatheter aortic valve implantation (TAVI) involves implanting a new valve within a previously implanted one.
- Balloon-expandable transcatheter heart valves (THVs) carry a risk of underexpansion when used in redo-TAVI procedures.
- Underexpansion can negatively impact THV performance and patient outcomes.
Purpose of the Study:
- To evaluate the effect of pre- and post-dilatation on the expansion of balloon-expandable THVs during redo-TAVI.
- To compare the expansion of SAPIEN 3 THVs implanted within SAPIEN XT or SAPIEN 3 failed valves.
Main Methods:
- Benchtop redo-TAVI simulations using 23 mm SAPIEN XT and SAPIEN 3 as index valves.
- Implantation of a 23 mm SAPIEN 3 THV within the index valves.
- Application of pre- and/or post-dilatation using a 23 mm non-compliant balloon.
- Assessment of THV expansion via microcomputed tomography (micro-CT) and hydrodynamic testing.
Main Results:
- Without dilatation, the redo-SAPIEN 3 THV was significantly underexpanded within both SAPIEN XT and SAPIEN 3.
- Pre- and post-dilatation improved redo-THV expansion, with an additive effect.
- Only SAPIEN 3 in SAPIEN 3 with both pre- and post-dilatation achieved nominal expansion.
- Underexpanded valves showed worse leaflet pinwheeling, despite acceptable hydrodynamic performance.
Conclusions:
- Full expansion of a 23 mm SAPIEN 3 THV in a redo-TAVI procedure is achievable within a SAPIEN 3 valve using both pre- and post-dilatation.
- SAPIEN 3 THVs implanted within SAPIEN XT valves remained underexpanded even with dilatation.
- CT assessment of THV expansion and the judicious use of pre-/post-dilatation are crucial in redo-TAVI.