Frame-rate reduction to reduce radiation dose for cardiac device implantation is safe
Fabian T Bork1, Andreas A Boehmer1, Celine Zezyk1
1Department of Cardiology, St. Josefs-Hospital Wiesbaden, Wiesbaden, Germany.
Background:
Radiation exposure to patient and surgeon during cardiac implantable electrical device (CIED) procedures remains a substantial health hazard to date. Advanced technical options for radiation dose reduction often pose considerable financial hurdles. We propose a near-zero cost, low-effort modification to a clinical x-ray system significantly reducing radiation dose during CIED implantation.
Objective:
We aim to evaluate a reduced frame rate protocol in CIED implantation for complication rates and reduction in radiation exposure.
Methods:
Starting May 2019, the frame rate during CIED implantations at our hospital was halved from 7.5 frames/s to 3.8 frames/s, and no further technical changes were made. During the following year, 264 patients were operated using this protocol and retrospectively compared with 231 cases implanted in the year before the protocol change, totaling 495 cases. Of these, 17%, 63%, and 19% were single-chamber, dual-chamber, or resynchronization devices, respectively. Incidence of complication prior to hospital discharge was considered the primary endpoint of the analysis. Radiation dose and procedural parameters were secondary endpoints.
Results:
There was no increase in complications with the reduced frame rate protocol. Regression analysis further supported that the reduced frame rate radiation protocol was not associated with complication rates. Radiation exposure measured as dose area product was significantly reduced by ∼62% (median 369 [interquartile range 154-1207] cGy·cm2 via the reduced frame rate protocol vs median 970 [interquartile range 400-1906] cGy·cm2 with the standard frame rate; P < 0.01).
Conclusion:
A reduction of frame rate during CIED implantation is safe in terms of complication incidence and effective in terms of reducing radiation exposure.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...