Effects of CT-based attenuation correction on quantitative SPECT/CT of jawbone
Tatsuya Tsuchitani1, Kazuhiro Kitajima, Yoshiyuki Takahashi
1Department of Radiological Technology, Hyogo College of Medicine College Hospital, Hyogo, Japan. kazu10041976@yahoo.co.jp.
Objective:
With single photon emission computed tomography (SPECT)/computed tomography (CT) quantitative examinations, CT-based attenuation correction (CTAC) is considered necessary, though its effect on the quantitative values of an examined area close to the body surface, such as the jawbone, has not been elucidated. We performed an investigation to determine whether quantitative evaluation using a bone SPECT standalone device without CT is possible.
Subjects And Methods:
The calculated indices were maximum standardized uptake value (SUVmax) and SUVpeak. Grouping was performed based on the presence or absence of CTAC. The CTAC group underwent CTAC, while the noAC group did not.Validation was performed using clinical data of patients who underwent a jawbone SPECT/CT examination. Becquerel calibration factor (BCF) is required for calculation of SUV, and was determined with values obtained with both phantom and syringe methods. The index for the uptake areas in each group was assessed using a paired t-test.
Results:
Using BCF obtained with the phantom method, both SUVmax and SUVpeak were higher in the noAC group. In contrast, BCF obtained with the syringe method showed no significant difference between the CTAC and noAC groups in regard to SUVmax and SUVpeak. This tendency was found regardless of the device used. Also, a high correlation was observed between the groups for both devices (r=0.95 and 0.93).
Conclusion:
Our findings show that BCF obtained with a syringe method should be used when performing quantitative evaluation without CTAC. They also indicate that quantitative evaluation using a SPECT standalone device may be possible for jawbone SPECT/CT examinations.
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies III: Computed Tomography
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...