Dynamic Ultrasound Focusing and Centimeter-Scale Ex Vivo Tissue Ablations With a CMUT Probe Developed for
Abstract:
Thermal ablation of localized prostate tumors via endocavitary ultrasound-guided high-intensity focused ultrasound (USgHIFU) faces challenges that could be alleviated by better integration of dual modalities (imaging/therapy). Capacitive micromachined ultrasound transducers (CMUTs) may provide an alternative to existing piezoelectric technologies by exhibiting advanced integration capability through miniaturization, broad frequency bandwidth, and potential for high electroacoustic efficiency. An endocavitary dual-mode USgHIFU probe was built to investigate the potential of using CMUT technologies for transrectal prostate cancer ablative therapy. The USgHIFU probe included a planar 64-element annular high-intensity focused ultrasound (HIFU) CMUT array ( [Formula: see text] = 3 MHz) surrounding a 256-element linear imaging CMUT array. Acoustic characterization of the HIFU array included 3-D pressure field mapping and radiation force balance measurements. Ex vivo proof-of-concept experiments consisted in generating HIFU thermal ablations with the CMUT probe on porcine liver tissues. The planar CMUT probe enabled HIFU dynamic focusing (distance range: 32-72 mm) while providing acoustic surface intensities of 1 W/cm2 that allowed producing elementary ex vivo ablations in depth of liver tissue ( L ×W ≈ 10×5 mm). Combinations of dynamic focusing, along with probe rotation and translation produced larger thermal ablations ( L ×W ≈ 20×20 mm) by juxtaposing multiple elementary ablations, consistent with expected results obtained through numerical modeling. The technical feasibility of using a USgHIFU probe, fully developed using CMUTs for tissue ablation purposes, was demonstrated. The HIFU-CMUT array showed tissue ablation capabilities with volumes compatible with localized cancer targeting, thus providing assets for further development of focal therapies.
More Related Videos
05:56Evaluation of the Feasibility, Safety, and Accuracy of an Intraoperative High-intensity Focused Ultrasound Device for Treating Liver Metastases
Published on: January 9, 2019
08:08Evaluating Targeting Accuracy in the Focal Plane for an Ultrasound-guided High-intensity Focused Ultrasound Phased-array System
Published on: March 6, 2019
Related Concept Videos
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Ultrasonography
During an ultrasonography procedure, a handheld device called...