Activity standardization of 99mTc by digitizer-based 4πce(LS)-γ and 4πce(PC)-γ coincidence counting
Marcell P Takács1, Karsten Kossert1, Ole Nähle1
1Physikalisch-Technische Bundesanstalt (PTB), Bundesallee 100, 38116, Braunschweig, Germany.
Abstract:
An aqueous solution of 99mTc was standardized by means of 4πce(LS)-γ and 4πce(PC)-γ coincidence counting. The activity concentration was obtained by efficiency extrapolation relying on four different variation techniques. In both setups, fast digitizers were used for data acquisition. Detected events were recorded in list mode format, and their coincidence relationship was analyzed in the subsequent offline analysis. The combined result was used to participate in an international comparison using the Transfer Instrument of the International Reference System.