Dependence of ghost on the incident light angle into dichroic mirror
Shuo Wang1,2, Min Hu1,2, Beini Sun1,2
1Key Laboratory of Laser Life Science, Ministry of Education, College of Biophotonics, South China Normal University, Guangzhou, Guangdong, China.
Abstract:
The dichroic mirror (DM) is a key component in microscope. We found a ghost in the reflection channel of a dual-channel fluorescence microscope and studied the relationship between the ghost and the incidence angle into the DM. The DM emission surface reflection generated ghost if the is not . We analyzed the distance and intensity relationship between the ghost and the primary image, which is -dependent and was demonstrated by imaging live cells and a stage micrometer. The ghost can be eliminated by placing the DM between objective and tube lens, but not between tube lens and detector, ensuring that the incident light into the DM is approximately parallel. Furthermore, the transmitted light of the DM is shifted towards a longer wavelength with increasing . Collectively, microscopists must carefully optimize the when designing a microscope to avoid the ghost.
Related Concept Videos
Phase Contrast and Differential Interference Contrast Microscopy
In-phase-contrast microscopes, interference between light directly passing through a cell and light refracted by cellular components is used to create high-contrast, high-resolution images without staining. It is the oldest and simplest type of microscope that creates an image by altering the wavelengths of light rays passing through the specimen. Altered wavelength paths are created using an annular stop in the condenser. The annular stop produces a hollow cone of...
Total Internal Reflection Fluorescence Microscopy