Related Experiment Video
Updated: Jun 19, 2026

Four-Dimensional CT Analysis Using Sequential 3D-3D Registration
Published on: November 23, 2019
A new three-dimensional patient-specific cutting guide for opening wedge high tibial osteotomy based on ct scan:
Federica Rosso1, Roberto Rossi2, Philippe Neyret3,4
1Department of Orthopaedics and Traumatology, AO Ordine Mauriziano Hospital, Largo Turati 62, 10128, Turin, Italy. federica.rosso@yahoo.it.
Purpose:
The aim of this study was to evaluate the accuracy of a patient-specific cutting guide on both coronal and sagittal alignment compared to the pre-operative planning in OWHTO.
Methods:
Twelve OWHTO on 6 cadaveric specimens were performed by 3 experienced knee surgeons using patient-specific cutting guides based on 3D pre-operative planning. Since the specimens had no major deformities, a fixed correction of 6° on the left and 10° on the right legs were carried out to simulate different scenarios. A pre-operative and post-OWHTO 3D CT scans were performed, and images were superimposed using the dedicated 3D planning software to align their reference axes. A pre-operative planning was performed considering both Medial Proximal Tibial Angle (MPTA) and Posterior Tibial Slope (PTS), and a patient-specific cutting guide was produced. Planned and post-OWHTO MPTA and PTS were evaluated (mean and standard deviation), and Pearson's correlation coefficient was calculated to assess precision and accuracy of the whole treatment.
Results:
A mean correction of 6,1° (SD 1,9°) and 1,2° (SD 1°) was obtained respectively in the coronal plane (MPTA) and in the sagittal plane (PTS). The average difference between planned and post-OWHTO MPTA and PTS was respectively 1,2° (SD 0,6°) and 1,2° (SD 1°) in the sagittal plane (PTS). Pearson's correlation coefficient demonstrated a good accuracy of the treatment in both coronal and sagittal plane (respectively r=0,95 and r=0,86). No lateral hinge fractures were detected at the post-operative CT scan.
Conclusion:
OWTHO performed with the help of 3D patient specific cutting guide on cadaveric specimens demonstrated good accuracy and reliability in obtaining the planned correction. In vivo studies are necessary to confirm these results and evaluate cost-effectiveness of this system.
Level Of Evidence:
Level IV cadaveric study.