Interference of cardiac implantable electronic devices and computed tomography imaging in the current era with a
Akihito Ideishi1,2, Kenichiro Yamagata1, Tatsuya Nishii3
1Department of Cardiovascular Medicine National Cerebral and Cardiovascular Center Suita Japan.
Introduction:
Cardiac implantable electronic devices are used in patients with cardiac rhythm disorders. Computed tomography irradiation is not prohibited for patients with cardiac implantable electronic devices, despite adverse events being reported. Hence, appropriate preparation and knowledge are required before computed tomography irradiation can be carried out in these patients. Since there is limited knowledge or literature about the influence of computed tomography irradiation in cases with recent cardiac implantable electronic devices, we aimed to evaluate the adverse events and elucidate the necessary and sufficient safety measures associated with this therapy.
Methods And Results:
We placed cardiac implantable electronic devices on an anthropomorphic phantom model and observed their electrical activity in electrograms, while various protocols of computed tomography irradiation were implemented and adverse events evaluated. Oversensing with pauses of up to 3.2 s was observed in standard computed tomography protocols, but ventricular tachyarrhythmia or other clinically significant events could not be confirmed. Oversensing with pauses of up to 8.0 s was observed and ventricular tachyarrhythmia was detected in the maximum-dose protocols. However, treatments such as antitachycardia pacing or shock therapy for ventricular tachyarrhythmia were not observed because of their absence.
Conclusion:
Computed tomography irradiation for patients using cardiac implantable electronic devices is highly unlikely to cause clinically significant adverse events with the device settings and computed tomography protocols currently being used. Changing or monitoring the device settings routinely before computed tomography irradiation is not necessarily required for most patients.
More Related Videos
08:19Protocol for the Evaluation of MRI Artifacts Caused by Metal Implants to Assess the Suitability of Implants and the Vulnerability of Pulse Sequences
Published on: May 17, 2018
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System IV: CMRI
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies III: Computed Tomography