Related Experiment Video
Updated: Jul 19, 2025

Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
Validation of automated image co-registration integrated into in-house software for voxel-based internal dosimetry on
André Luiz Alberti Leitão1, Uysha de Souza Fonda2, Carlos Alberto Buchpiguel3
1Centers for Radiology and Nuclear Medicine - Centro Médico, Brasília, DF, Brazil.
Objective:
To develop an automated co-registration system and test its performance, with and without a fiducial marker, on single-photon emission computed tomography (SPECT) images.
Materials And Methods:
Three SPECT/CT scans were acquired for each rotation of a Jaszczak phantom (to 0°, 5°, and 10° in relation to the bed axis), with and without a fiducial marker. Two rigid co-registration software packages-SPM12 and NMDose-coreg-were employed, and the percent root mean square error (%RMSE) was calculated in order to assess the quality of the co-registrations. Uniformity, contrast, and resolution were measured before and after co-registration. The NMDose-coreg software was employed to calculate the renal doses in 12 patients treated with 177Lu-DOTATATE, and we compared those with the values obtained with the Organ Level INternal Dose Assessment for EXponential Modeling (OLINDA/EXM) software.
Results:
The use of a fiducial marker had no significant effect on the quality of co-registration on SPECT images, as measured by %RMSE (p = 0.40). After co-registration, uniformity, contrast, and resolution did not differ between the images acquired with fiducial markers and those acquired without. Preliminary clinical application showed mean total processing times of 9 ± 3 min/patient for NMDose-coreg and 64 ± 10 min/patient for OLINDA/EXM, with a strong correlation between the two, despite the lower renal doses obtained with NMDose-coreg.
Conclusion:
The use of NMDose-coreg allows fast co-registration of SPECT images, with no loss of uniformity, contrast, or resolution. The use of a fiducial marker does not appear to increase the accuracy of co-registration on phantoms.
More Related Videos
09:36In vivo Positron Emission Tomography to Reveal Activity Patterns Induced by Deep Brain Stimulation in Rats
Published on: March 23, 2022
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...