Technical Note: Assessment of Pulse Pileup on Single-Energy and Multienergy Images From a Clinical Photon-Counting
Joshua Ray Chen1, Timothy N Winfree, Michael R Bruesewitz
1From the Department of Radiology, Mayo Clinic, Rochester, MN.
Objective:
Pulse pileup effects occur when pulses occur so close together that they fall on top of one another, resulting in count loss and errors in energy thresholding. To date, there has been little work systematically detailing the quantitative effects of pulse pileup on material decomposition accuracy for photon-counting detector (PCD) computed tomography (CT). Our aim in this work was to quantify the effects of pulse pileup on single-energy and multienergy CT images, including low-energy bin (BL), high-energy bin (BH), iodine map, and virtual noncontrast images from a commercial PCD-CT.
Methods:
Scans of a 20-cm diameter multienergy CT phantom with 10 solid inserts were acquired at a fixed tube potential as the tube current was varied across the available range. Four types of images (BL, BH, iodine map, and virtual noncontrast) were reconstructed using an iterative reconstruction algorithm at strength 2, a quantitative reconstruction kernel (Qr40), 2-/1-mm slice thickness/increment, and a 210-mm field-of-view. The mean and standard deviation of CT numbers were recorded and the ratios of CT number between BL and BH images were calculated and plotted, along with noise versus tube current and noise × versus tube current.
Results:
As tube current was increased, the range of variations in CT numbers was less than 13.4 HU for all inserts and image types evaluated. Noise × versus tube current showed a small positive slope equal to a noise increase from 100 mA of 10% at 500 mA and 15% at 900 mA compared with what would be expected if the slope was zero.
Conclusions:
Minimal impact on single-energy and multienergy CT numbers and noise performance was observed for the evaluated clinical PCD-CT system.
More Related Videos
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies III: Computed Tomography
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...