Ultrasound Cyclo Plasty in Eyes with Glaucoma
Published on: January 26, 2018
XEN®-63 Compared to XEN®-45 Gel Stents to Reduce Intraocular Pressure in Glaucoma
Charlotte Evers1, Daniel Böhringer1, Sara Kallee1
1Eye Center, Medical Center-University of Freiburg, Faculty of Medicine, University of Freiburg, 79106 Freiburg, Germany.
Abstract:
The XEN® gel stent reduces intraocular pressure (IOP) in glaucoma. XEN®-45 is widely used; the newer XEN®-63 has a larger lumen targeting potentially lower IOP outcomes. We retrospectively compared the first 15 XEN®-63 cases to 15 matched XEN®-45 controls. With a preoperative IOP of 18.1 ± 3.9 mmHg (mean ± SD) and a final IOP of 9.1 ± 2.0 mmHg, XEN®-63 implantation resulted in an IOP reduction of 44.6 ± 16.5%. Similarly, with a preoperative IOP of 18.3 ± 4.5 mmHg and a final IOP of 10.3 ± 2.1 mmHg, XEN®-45 implantation resulted in an IOP reduction of 40.1 ± 17.2%. The median follow-up period was 204 days (range 78-338 days) for the XEN®-63 group and 386 days (range 99-1688 days) for the XEN®-45 group. In total, 5/15 eyes of each group underwent open conjunctival bleb revision within the period of observation. Three eyes of the XEN®-63 group had secondary glaucoma surgery. One eye in the XEN®-63 group and three eyes in the XEN®-45 group required a restart of antiglaucomatous medication. In conclusion, both stents effectively lower IOP and medication. XEN®-63 achieved a slightly lower IOP over a short follow-up. Complication and revision rates were similar.