Miniaturization and Automation Protocol of a Urinary Organic Acid Liquid-Liquid Extraction Method on GC-MS
Masauso Moses Phiri1,2, Elmarie Davoren2, Barend Christiaan Vorster2
1Department of Pathology and Microbiology, School of Medicine, University of Zambia, Lusaka 10101, Zambia.
Abstract:
The aim of this study was to improve the extraction method for urinary organic acids by miniaturizing and automating the process. Currently, manual extraction methods are commonly used, which can be time-consuming and lead to variations in test results. To address these issues, we reassessed and miniaturized the in-house extraction method, reducing the number of steps and the sample-to-solvent volumes required. The evaluated miniaturized method was translated into an automated extraction procedure on a MicroLab (ML) Star (Hamilton Technologies) liquid handler. This was then validated using samples obtained from the ERNDIM External Quality Assurance program. The organic acid extraction method was successfully miniaturized and automated using the Autosampler robot. The linear range for most of the thirteen standard analytes fell between 0 to 300 mg/L in spiked synthetic urine, with low (50 mg/L), medium (100 mg/L), and high (500 mg/L) levels. The correlation coefficient (r) for most analytes was >0.99, indicating a strong relationship between the measured values. Furthermore, the automated extraction method demonstrated acceptable precision, as most organic acids had coefficients of variation (CVs) below 20%. In conclusion, the automated extraction method provided comparable or even superior results compared to the current in-house method. It has the potential to reduce solvent volumes used during extraction, increase sample throughput, and minimize variability and random errors in routine diagnostic settings.
More Related Videos
09:38Single-throughput Complementary High-resolution Analytical Techniques for Characterizing Complex Natural Organic Matter Mixtures
Published on: January 7, 2019
08:41Integration of Miniaturized Solid Phase Extraction and LC-MS/MS Detection of 3-Nitrotyrosine in Human Urine for Clinical Applications
Published on: July 14, 2017
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall....
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...