Size based dependence of patient dose metrics, and image quality metrics for clinical indicator-based imaging
I Sebelego1, S Acho2, B van der Merwe1
1Department of Clinical Sciences, Faculty of Health and Environmental Sciences, Central University of Technology, Bloemfontein, South Africa.
Introduction:
Diagnostic reference level (DRL) values for computed tomography (CT) based on clinical indication are warranted since imaging protocols are indication-dependent. This study proposes clinical DRL values using the CT dose metrics and five patient size-related parameters while considering image quality.
Methods:
The volumetric CT dose index (CTDIvol), dose-length product (DLP) and five size-related parameters of size-specific dose estimates (SSDE), namely the anterior-posterior (AP) dimension, lateral (LAT) dimension, sum dimension, effective diameter, and the body mass index (BMI), were used to calculate DRL values for CT chest-abdomen-pelvis (CAP) and abdomen-pelvis (AbP) protocols. DRL values of the clinical indications for cancer, urinary system stones and other pathologies were assessed based on the BMI classifications using the median and 75th percentile. An image subtraction algorithm was used to assess the image quality metrics (IQM) of the CT images.
Results:
The 75th percentile for SSDEAP dimension for CAP cancer was 19.7, 14.9 and 12.7 mGy at Hospitals A, C and E, respectively. The median DLP for other AbP pathologies was 556.3, 1452.0 and 1960.7 mGy.cm for normal weight, overweight and obese patients, respectively, at Hospital A. The image quality varied among BMI classifications for different clinically indicated examinations. Although the dose increased with BMI, the image quality index was consistent because automatic tube current modulation (ATCM) was used.
Conclusion:
DRL values are influenced by patient size-related parameters and the clinical indication protocols, while the image quality index is independent of the BMI.
Implications For Practice:
Size-related clinical DRL values and image quality index can be used to monitor and optimise dose and image quality. Acquisition parameters and image quality indexes should be investigated and adjusted when unusually high DRL values are noted.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
Related Concept Videos
Radiological Investigation I: X-ray and CT
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies III: Computed Tomography
Imaging Studies for Cardiovascular System V: CT
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...