Compared with physician overread, computer is less accurate but helpful in interpretation of electrocardiography for
Yuan-Hui Wu1, Ai-Hsien Li2, Tsan-Chi Chen3
1Department of Emergency Medicine, Far Eastern Memorial Hospital, New Taipei City, Taiwan; School of Medicine, Fu-Jen Catholic University, New Taipei City, Taiwan.
Introduction:
Previous studies have demonstrated varying sensitivity and specificity of computer-interpreted electrocardiography (CIE) in identifying ST-segment elevation myocardial infarction (STEMI). This study aims to evaluate the accuracy of contemporary computer software in recognizing electrocardiography (ECG) signs characteristic of STEMI compared to emergency physician overread in clinical practice.
Material And Methods:
In this retrospective observational single-center study, we reviewed the records of patients in the emergency department (ED) who underwent ECGs and troponin tests. Both the Philips DXL 16-Lead ECG. Algorithm and on-duty emergency physicians interpreted each standard 12‑lead ECG. The sensitivity and specificity of computer interpretation and physician overread ECGs for the definite diagnosis of STEMI were calculated and compared.
Results:
Among the 9340 patients included in the final analysis, 133 were definitively diagnosed with STEMI. When "computer-reported infarct or injury" was used as the indicator, the sensitivity was 87.2% (95% CI 80.3% to 92.4%) and the specificity was 86.2% (95% CI 85.5% to 86.9%). When "physician-overread STEMI" was used as the indicator, the sensitivity was 88.0% (95% CI 81.2% to 93.0%) and the specificity was 99.9% (95% CI 99.8% to 99.9%). The area under the receiver operating characteristic curve for physician-overread STEMI and computer-reported infarct or injury were 0.939 (95% CI 0.907 to 0.972) and 0.867 (95% CI 0.834 to 0.900), respectively.
Conclusions:
This study reveals that while the sensitivity of the computer in recognizing ECG signs of STEMI is similar to that of physicians, physician overread of ECGs is more specific and, therefore, more accurate than CIE.
More Related Videos
Related Concept Videos
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Acute Coronary Syndrome III: Diagnostic Studies
ECG Interpretation of Rhythms
Components of the Electrocardiogram
The primary components of a normal ECG waveform in Normal sinus rhythm(NSR) include the P wave, PR interval, QRS complex, ST segment, T wave, and occasionally a U wave.
ECG waveforms are divided by vertical and horizontal lines at standard intervals.
The horizontal axis measures time and rate, and the vertical axis measures amplitude or voltage....
Acute Coronary Syndrome I: Introduction
Acute Coronary Syndrome II: Pathophysiology and Clinical Manifestations
Exercise Stress Test
Exercise stress testing, commonly known as a treadmill test, is a noninvasive procedure used to evaluate cardiovascular function and diagnose heart conditions.
Definition
An exercise stress test measures the heart's response to exertion using a treadmill or stationary bicycle. Chest electrodes record the heart's electrical activity through an ECG, and blood pressure is monitored regularly.
Purposes