Applicability evaluation of the TRS-483 protocol for the determination of small-field output factors using different
Kohei Kawata1, Hideaki Hirashima2, Yusuke Tsuruta3
1Division of Clinical Radiology Service, Kyoto University Hospital, Kyoto, Japan.
Purpose:
To evaluate the applicability of TRS-483 output correction factors (CFs) for small-field output factors (OFs) using different multi-leaf collimators (MLC) and field-shaping types.
Methods:
All measurements were performed on TrueBeam, TrueBeam STx, and Halcyon using 6 MV flattening filter-free energy. Four detectors, including CC01, CC04, microDiamond, and EDGE, were used. Nominal field sizes ranging from 1 × 1 to 4 × 4, and 10 × 10 cm2 were used to measure small-field OFs at source-to-axis distance of 100 cm with a 0° gantry angle in a 3D water phantom. Further, the field-shaping types were defined using jaw collimator or MLC (five different configurations). A field size of 10 × 10 cm2 was used as the reference for calculation of OFs obtained as ratio of detector readings (OFdet). The percentage difference and coefficient of variation of OFdet and OFdet corrected by applying CF were compared for each field size and configuration.
Results:
For OFdet corrected by applying CF, the ranges of percentage difference and coefficient of variation in all configurations for ≥ 2 × 2 cm2 fields were reduced from 1.2-2.2 to 0.8-1.3 percentage points (%pt) and from 0.5-1.0 to 0.4-0.7%, respectively. For 1 × 1 cm2 field, the ranges of percentage difference and coefficient of variation were reduced from 3.3-5.7 to 1.2-2.2 %pt and from 2.2-3.7 to 0.8-1.1%, respectively.
Conclusions:
The CFs described in TRS-483 dosimetry protocol have broad applicability in reducing OF variations between detectors under different MLC and field-shaping types.