Transient conduction disturbances acutely after pulsed-field cavotricuspid isthmus ablation: a case report
George Andrikopoulos1, Konstantinos Tampakis1, Alexandros Sykiotis1
1First Department of Cardiology/Electrophysiology and Pacing, Henry Dunant Hospital Center, 107 Mesogeion ave, 11526 Athens, Greece.
Background:
Cavotricuspid isthmus pulsed-field ablation has been recently described to be safely performed despite initial reports on coronary arterial spasm while conduction disturbances as a complication of cavotricuspid isthmus ablation are rare and have been reported exclusively for radiofrequency catheter ablation.
Case Summary:
A 64-year-old female patient with mechanical prosthetic valves underwent atrial fibrillation ablation using the pentaspline pulsed-field ablation catheter. At the end of the uneventful pulmonary vein isolation, an atrial tachycardia depended to the cavotricuspid isthmus occurred. A single pulsed-field application at the cavotricuspid isthmus resulted in right bundle branch block combined with posterior fascicular hemiblock and PR prolongation that resolved spontaneously within 12 h.
Discussion:
This is the first report of transient conduction disturbances as a complication of cavotricuspid isthmus pulsed-field ablation. Although the underlying mechanism, either single or miscellaneous, was not verified, this case highlights that caution should be taken when the pentaspline pulsed-field ablation catheter is used for cavotricuspid isthmus ablation.