A Comparison of Proximal and Tracheal Airway Pressures During Pressure Controlled Ventilation
Mark O Zander1, Nikola Stankovic2, Mirko Meboldt1
1Product Development Group Zurich, Department of Mechanical and Process Engineering, ETH Zurich, Zurich, Switzerland.
Background:
Airway pressure is usually measured by sensors placed in the ventilator or on the ventilator side of the endotracheal tube (ETT), at the Y-piece. These remote measurements serve as a surrogate for the tracheal or alveolar pressure. Tracheal pressure can only be predicted correctly by using a model that incorporates the pressure at the remote location, the flow through the ETT, and the resistance of the ETT if the latter is a predictable function of Y-piece flow. However, this is not consistently appropriate, and accuracy of prediction is hampered.
Methods:
This in vitro study systematically examined the ventilator pressure in dependence of compliance of the respiratory system (CRS), inspiratory time, and expiratory time during pressure-controlled ventilation by using a small intratracheal pressure sensor and a mechanical lung simulator. Pressures were measured simultaneously at the ventilator outlet, at the Y-piece, and in the trachea during pressure-controlled ventilation with a peak inspiratory pressure of 20 cm H2O and a PEEP of 5 cm H2O while changing CRS (10, 30, 60, 90, and 100 mL/cm H2O) and varying inspiratory time and expiratory time.
Results:
Tracheal pressures were always lower (maximum 8 cm H2O during inspiration) or higher (maximum 4 cm H2O during expiration) than the pressures measured proximal to the ETT if zero-flow conditions were not achieved at the end of the breathing cycles.
Conclusions:
Dependent on CRS and the breathing cycle, tracheal pressures deviated from those measured proximal to the ETT under non-zero-flow conditions. Intratracheal pressure and pressure curve dynamics can differ greatly from the ventilator pressure, depending on the ventilator setting and the CRS. The small pressure sensor may be used as a measurement method of tracheal pressure via integration onto an ETT.
Related Concept Videos
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Pressure Relationships in Thoracic Cavity
Breathing Mechanisms
Both intra-alveolar and intrapleural pressures rely on specific lung properties. The ability to breathe—allowing air to enter the lungs...
Ventilatory Modes
There are three ventilatory modes: full support, partial support, and spontaneous. These are described below.
Full Support Modes
Full support modes include controlled mechanical ventilation, continuous mandatory...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Pulmonary Ventilation: Inhalation
Boyle's law becomes particularly pertinent when examining respiratory...
Mechanical Ventilation I: Indication and Settings