Initial Experience With Intercostal Insertion of an Extravascular ICD Lead Compatible With Existing Pulse Generators
Martin C Burke1,2,3, Reinoud E Knops2, Vivek Reddy4
1CorVita Science Foundation, Chicago, IL (M.C.B.).
Background:
This study assessed safety and feasibility of a novel extravascular implantable cardioverter defibrillator (ICD) lead when inserted anteriorly through a rib space and connected to various commercially available ICD pulse generators (PGs) placed in either a left mid-axillary or left pectoral pocket. Currently available or investigational, extravascular-ICDs include a subcutaneous or subxiphoid lead connected to customized extravascular-ICD PGs.
Methods:
This novel extravascular-ICD (AtaCor Medical Inc, San Clemente, CA) employs a unique intercostal implant technique and is designed to function with commercial DF-4 ICD PGs. In this nonrandomized, single-center, acute study, 36 de novo or replacement ICD (transvenous ICD) patients enrolled to receive a concomitant extravascular-ICD lead inserted through an intercostal space along the left parasternal margin. extravascular-ICD leads were connected to DF-4 compatible ICD PGs positioned in either a left mid-axillary or pectoral pocket for acute sensing and defibrillation testing. Defibrillation testing started at 30 Joules (J) and stepped up or down in 5 to 10 joule increments depending on the success and limitations of the generator used.
Results:
Successful acute defibrillation using ≤35 J was noted in 100% of left mid-axillary PG subjects (n=27, mean 16.3±8.6 J) and 83% of left pectoral PG subjects (n=6, mean 21.0±8.4 J). Furthermore, 24 of 27 (89%) of patients tested with a left, mid-axillary intermuscular PG had successful VF conversion with defibrillation energies at least 10 J below the maximum delivered output of the device. All evaluable episodes (n=93) were automatically sensed, detected, and shocked. No serious device-related intraoperative adverse events were observed.
Conclusions:
This first-in-human study documented the safe and reliable placement of a novel extravascular ICD lead with effective sensing and defibrillation of induced ventricular fibrillation using commercial DF-4 ICD PGs.
More Related Videos
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
07:41Insertion, Maintenance, and Removal of the Percutaneous Dual Lumen Cannula Right Ventricular Assist Device
Published on: July 20, 2022
Related Concept Videos
Cardiac Catheterization I: Pre-Procedure Overview
Cardiac Catheterization II: Right Heart Catheterization
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization IV: Nursing Management
Cardiopulmonary Resuscitation III: AED Use
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...