Related Experiment Video
Updated: Jul 19, 2025

Dynamic Multiparameter Platelet Function Assessment Using a Capacitive Biosensor
Published on: May 2, 2025
Metal-pad-enhanced resistive pulse sensor reveals complex-valued Braess paradox
Alan Dong1, Lydia Sohn2, Michael Lustig1
1Department of Electrical Engineering and Computer Sciences, University of California, Berkeley, Berkeley, California 94720, USA.
Abstract:
A resistive pulse sensor measures the electrical impedance of an electrolyte-filled channel as particles flow through it. Ordinarily, the presence of a nonconductive particle increases the impedance of the channel. Here we report a surprising experimental result in which a microfluidic resistive pulse sensor experiences the opposite effect: The presence of a nonconductive particle decreases the channel impedance. We explain the counterintuitive phenomenon by relating to the Braess paradox from traffic network theory, and we call it the complex-valued Braess paradox (CVBP). We develop theoretical models to study the CVBP and corroborate the experimental data using finite element simulations and lumped-element circuit modeling. We then discuss implications and potential applications of the CVBP in resistive pulse sensing and beyond.
Related Concept Videos
Measurement of Blood Pressure
Resistance and Conductance
Various factors impact the resistance of a conductor. Spiraling in stranded conductors increases their...
Pulse
The pulse serves as a clinical...
Resting Membrane Potential
The Inside of a Neuron is More Negative
The membrane potential of a cell can be measured by inserting a microelectrode into a cell and comparing the charge to a reference electrode in the extracellular fluid. The...