Heat transfer analysis study for the design of the New Polychromatic Beam Neutron Reflectometer CANDOR (Chromatic
Lior Uzan1,2, Juscelino B Leão2, Christoph Brocker2
1Nuclear Research Center Negev, Israel.
Abstract:
A newly developed polychromatic beam neutron reflectometer CANDOR (Chromatic Analysis Neutron Diffractometer Or Reflectometer) on NG-1 at the NIST Center for Neutron research (NCNR) utilizes a wavelength-sensitive neutron detector consisting of 324 analyzing highly-oriented pyrolytic graphite (HOPG) crystals positioned sequentially in rows. Known for having a small thermal diffuse scattering cross section, HOPG crystals can lead to low signal-to-noise ratios in wavelength-sensitive detectors such as CANDOR. Even though it is possible to mathematically separate the desired signal from thermal diffuse scattering; by cooling the detector array of HOPG crystals in order to minimize the Debye Waller effect generates a better solution to this problem. In this heat transfer analysis study we show, within the instrument design constrains and thermodynamic considerations, technical feasibility and test results for the development of the New Polychromatic Beam Neutron Reflectometer CANDOR (Chromatic Analysis Neutron Diffractometer Or Reflectometer) at the NIST Center for Neutron Research.
More Related Videos
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021
09:23Author Spotlight: Advancing Energy Solutions Using Nanocomposites as Processed Thermoelectric Materials
Published on: May 17, 2024