Determination of serum CA724 levels using fluorescence immunochromatography
Related Experiment Video
Updated: Jul 18, 2025

Standardization of a Cytometric Bead Assay Based on Egg-Yolk Antibodies
Published on: May 19, 2023
Chang Liu1, Cuicui Chen2,3, Fenglan Peng4
1Heilongjiang Academy of Chinese Medicine Sciences, Harbin, 150036, China.
Background:
Carbohydrate antigen 724 (CA724) is a sensitive and specific indicator for multiple malignant tumors. The aim of this study was to establish a Eu-time resolved fluorescence immunochromatography (Eu-TRFICO) method for quantitative detection of CA724 in serum.
Methods:
Eu-TRFICO strips were optimized and assembled. The sensitivity, specificity and precision were evaluated using CA724 standard dilutions and matrix serum. Meanwhile, the reference interval, comparison, and sensitivity/specificity were performed using clinical negative/positive gastric cancer serum samples.
Results:
The standard curve equation was y = 9.869 x - 154.12 (R2 = 0.993), and the sensitivity was 0.42 U/mL. The common interferents in serum could not affect the quantitative results with low cross-reactivities (all no more than 1.09%). All average recoveries of the intra- and interbatch ranged from 102.38 to 106.40%, and all CVs were below 10%. The reference interval of the healthy subjects was < 4.68 U/mL and the reference interval of the subjects with grade I/II gastric cancer was > 9.54 U/mL. Additionally, a high Pearson r (0.9503) and sensitivity/specificity (92.86%/94.20%) were obtained.
Conclusion:
This study prepared Eu-TRFICO strips with high sensitivity, specificity, precision and satisfactory clinical testing performance, which provides more options for clinical quantitative and convenient testing of CA724.