Performing Intracochlear Electrocochleography During Cochlear Implantation
Published on: March 8, 2022
Effect of congenital inner ear malformations (IEMs) on electrically evoked compound action potential (ECAP) responses
Nithreen M Said1,2, Lena S Telmesani3, Laila M Telmesani3
1Audiovestibular Medicine Unit, Otorhinolaryngology Department, Faculty of Medicine, Imam Abdulrahman Bin Faisal University, Dammam, Saudi Arabia. nithreens@yahoo.com.
Purpose:
The study was designed to assess the electrically evoked compound action potential (ECAP) responses in children with inner ear malformations compared to children with normal inner ear anatomy.
Methods:
The study included 235 prelingual deaf children who were implanted in cochlear implant unit in King Fahad University hospital-Imam Abdulrahman Bin Faisel University. Subjects were using either Cochlear Nucleus or Medel cochlear implant devices. We had 171 (64.5%) subjects with normal inner ear anatomy and 94 (35.5%) subjects with inner ear malformations (IEMs) and they were classified into 6 groups according to inner ear anatomy. Fourteen subjects (14.9%) subjects had enlarged vestibular aqueduct (EVA), 30 (32%) subjects had Mondini deformity, 25 (26.6%) subjects had incomplete partition type two (IPII), 9 (9.6%) subjects had incomplete partition type one (IPI) and 16 (17%) subjects had hypoplastic cochlea type III or IV. Intraoperative electrically evoked compound action potential (ECAP) responses were analyzed and compared in all subjects.
Results And Conclusions:
Measurable ECAP responses can be elicited in patients with IEMs in most of the channels. Severe malformations can affect the prevalence of measuring ECAP and getting identifiable waveform morphology. Additionally, increased thresholds and lower slope of AGF was observed in IEMs specially in more severe malformations (e.g. IPI). IPI patients with better word recognition scores tended to show more identifiable ECAP measurements. This could suggest the presence of some correlation between ECAP responses and patients' performance after cochlear implantation.
More Related Videos
09:06Cochlear Implant Surgery and Electrically-evoked Auditory Brainstem Response Recordings in C57BL/6 Mice
Published on: January 9, 2019
07:00Electrically Evoked Stapedius Reflex Measurements in Cochlear Implantation and Its Application in the Postoperative Fitting Process
Published on: June 21, 2024
Related Concept Videos
The Cochlea
Hair Cells
Auditory Pathway
When viewed cross-sectionally, the cochlea reveals the scala vestibuli and scala tympani flanking...