Related Experiment Video
Updated: Jul 18, 2025

Use of a Percutaneous Ventricular Assist Device/Left Atrium to Femoral Artery Bypass System for Cardiogenic Shock
Published on: August 16, 2021
Short-term percutaneous mechanical circulatory support: no promise without positioning!
Tim Balthazar1,2, Nicolas M Van Mieghem3, Matthias Raes2
1Centre for Cardiovascular Diseases, University Hospital Brussels, Laarbeeklaan 101, Jette 1090, Belgium.
Abstract:
Short-term percutaneous mechanical circulatory support by a micro-axial flow pump is increasingly used to support the left ventricle in cardiogenic shock. After a correct indication and placement, appropriate device management in the cardiac intensive care unit is vital to ensure optimal pump function and adequate haemodynamic support. A key element hereby is a correct percutaneous ventricular assist device (pVAD) position. This review explains how an optimal left-sided pVAD position can be achieved and maintained, focusing on the correct insertion depth and rotational angle. Useful imaging techniques, placement and replacement manoeuvres, and monitoring options through the console are discussed. The frequently encountered problem of mal-rotation towards the mitral valve, which may cause suction alarms, haemolysis, aortic regurgitation, and inadequate haemodynamic support, is explained. Finally, a practical bedside approach to assess pVAD position and discern suction alarms due to mal-positioning from haemodynamic problems is proposed.