Background:
Jet nebulizers are commonly used for bronchodilator therapy in COPD. High-flow nasal cannula with vibrating mesh nebulizer (HFNC-VMN) is a recently developed system; however, few studies have compared the efficacy of bronchodilator administration via HFNC-VMN to jet nebulizer in stable COPD. This study aimed to compare the effect of salbutamol administered via HFNC-VMN versus jet nebulizer on airway and lung function in subjects with stable COPD.
Methods:
This randomized non-inferiority crossover physiologic study enrolled subjects with stable COPD. Salbutamol was nebulized via HFNC-VMN or jet nebulizer in random order with a 4-h washout period between crossover sequences. Spirometry, lung volume, and impulse oscillometry were performed at baseline and after each intervention. The primary outcome was change in FEV1 from baseline. Secondary outcomes included changes in other respiratory-related parameters and nebulization time compared between the 2 devices.
Results:
Seventeen subjects were enrolled. HFNC-VMN and jet nebulizer both significantly improved FEV1 from baseline (P = .005 and P = .002, respectively). The difference between respiratory resistance at 5 Hz and 20 Hz significantly decreased after HFNC-VMN compared to baseline (P = .02), while no significant change was observed after jet nebulizer (P = .056). Area of reactance and resonant frequency of reactance were both significantly decreased (P = .035 and P = .03, respectively), and respiratory reactance at 5 Hz significantly increased (P = .02) in the HFNC-VMN group compared to baseline indicating improved lung mechanics, with no significant changes with the jet nebulizer. HFNC-VMN had a shorter nebulization time (6 [5-9] min vs 20 [16-22] min, respectively, P < .001).
Conclusions:
Bronchodilator therapy via HFNC-VMN was not inferior to jet nebulizer for subjects with stable COPD and can significantly improve airway oscillometry mechanics and decrease nebulization time compared to jet nebulizer.
Related Concept Videos
Inhaled Medications
COPD: Management Using Bronchodilators and Corticosteroids
Administering Oxygen by Nasal Cannula
Nasal Cannulas
A nasal cannula is a lightweight tube split into two prongs placed in the nostrils,...
Mechanical Ventilation II: Invasive Ventilation
Negative-Pressure Ventilators
Negative-pressure ventilators create a vacuum around the chest or body to draw air into the lungs, simulating breathing. This method does not require an...
Mechanical Ventilation III: Noninvasive Ventilation
Noninvasive Positive-Pressure Ventilation...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...