Related Experiment Video
Updated: Jul 18, 2025

Pulse Wave Velocity Testing in the Baltimore Longitudinal Study of Aging
Published on: February 7, 2014
A Physical Model-Based Approach to One-Point Calibration of Pulse Transit Time to Blood Pressure
Objective:
To develop a novel physical model-based approach to enable 1-point calibration of pulse transit time (PTT) to blood pressure (BP).
Methods:
The proposed PTT-BP calibration model is derived by combining the Bramwell-Hill equation and a phenomenological model of the arterial compliance (AC) curve. By imposing a physiologically plausible constraint on the skewness of AC at positive and negative transmural pressures, the number of tunable parameters in the PTT-BP calibration model reduces to 1. Hence, as opposed to most existing PTT-BP calibration models requiring multiple (≥2) PTT-BP measurements to personalize, the PTT-BP calibration model can be personalized to an individual subject using a single PTT-BP measurement pair. Equipped with the physically relevant PTT-AC and AC-BP relationships, the proposed approach may serve as a universal means to calibrate PTT to BP over a wide BP range. The validity and proof-of-concept of the proposed approach were evaluated using PTT and BP measurements collected from 22 healthy young volunteers undergoing large BP changes.
Results:
The proposed approach modestly yet significantly outperformed an empiric linear PTT-BP calibration with a group-average slope and subject-specific intercept in terms of bias (5.5 mmHg vs 6.4 mmHg), precision (8.4 mmHg vs 9.4 mmHg), mean absolute error (7.8 mmHg vs 8.8 mmHg), and root-mean-squared error (8.7 mmHg vs 10.3 mmHg, all in the case of diastolic BP).
Conclusion:
We demonstrated the preliminary proof-of-concept of an innovative physical model-based approach to one-point PTT-BP calibration.
Significance:
The proposed physical model-based approach has the potential to enable more accurate and convenient calibration of PTT to BP.
More Related Videos
11:26Assessing Cerebral Autoregulation via Oscillatory Lower Body Negative Pressure and Projection Pursuit Regression
Published on: December 10, 2014
10:21Continuous Blood Sampling in Small Animal Positron Emission Tomography/Computed Tomography Enables the Measurement of the Arterial Input Function
Published on: August 8, 2019
Related Concept Videos
Measurement of Blood Pressure
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Assessment of blood pressure in brachial artery(two-step method)
Sites for measruring blood pressure
The Brachial Artery: Primary Site for Blood Pressure Measurement
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessment of blood pressure in brachial artery(one-step method)
Prepare for the Procedure: