Background:
The influence of power, duration and contact force (CF) on radiofrequency (RF) lesion formation is well known, whereas data on local impedance (LI) and electrode-tissue-coverage (ETC) is scarce. The objective was to investigate their effect on lesion formation in an ex vivo model.
Methods And Results:
An ex vivo model was developed utilizing cross-sections of porcine heart preparations and a force-sensing, LI-measuring catheter. N = 72 lesion were created systematically varying ETC (minor/full), CF (1-5 g, 10-15 g, 20-25 g) and power (20 W, 30 W, 40 W, 50 W). In minor ETC, the distal tip of the catheter was in electric contact with the tissue, in full ETC the whole catheter tip was embedded within the tissue. Lesion size and all parameters were measured once per second (n = 3320). LI correlated strongly with lesion depth (r = -0.742 for ΔLI; r = 0.781 for %LI-drop). Lesions in full ETC were significantly wider and deeper compared to minor ETC (p < .001) and steam pops were more likely. Baseline LI, ΔLI, and %LI-drop were significantly higher in full ETC (p < .001). In lesions resulting in steam pops, baseline LI, and ΔLI were significantly higher. The influence of CF on lesion size was higher in minor ETC than in full ETC.
Conclusions:
ETC is a main determinant of lesion size and occurrence of steam pops. Baseline LI and LI-drop are useful surrogate parameters for real-time assessment of ETC and ΔLI correlates strongly with lesion size.
More Related Videos
06:54Author Spotlight: Advancements in Impedance Monitoring for Cochlear Implant Surgery
Published on: August 4, 2023
08:15Fabrication of Fine Electrodes on the Tip of Hypodermic Needle Using Photoresist Spray Coating and Flexible Photomask for Biomedical Applications
Published on: November 28, 2017
Related Concept Videos
Bode Plots Construction
RLC Series Circuits: Impedance
Thus, the magnitude of the impedance is given by the following equation,
Mesh Analysis for AC Circuits
The process of harmonizing these impedances begins with a clear understanding of the input and output signals. Once these signals are known, the...