S Nithiananthan1, D Mirota2, A Uneri2
1Department of Biomedical Engineering, Johns Hopkins University, Baltimore, MD 21205.
Summary
This study introduces a novel deformable registration method for cone-beam CT (CBCT) guided surgery. The enhanced algorithm accurately accounts for excised tissue, improving intraoperative image guidance in head and neck procedures.
Area of Science:
- Medical Imaging
- Image Registration
- Surgical Navigation
Background:
- Intraoperative imaging, such as cone-beam CT (CBCT), is crucial for guiding head and neck surgery.
- Existing deformable registration methods struggle with anatomical changes due to tissue excision during surgery.
- Accurate registration is essential for real-time surgical adjustments and improved patient outcomes.
Purpose of the Study:
- To investigate the feasibility of fast, accurate deformable registration for intraoperative images with surgical excisions.
- To develop and evaluate a modified deformable registration algorithm capable of handling excised tissue.
- To enhance the precision of CBCT-guided head and neck surgical procedures.
Main Methods:
- Developed a novel deformable registration approach by incorporating an extra dimension to manage excised voxels.
- Utilized a modified Demons deformable registration algorithm to address limitations with excised tissue.
- Employed cadaveric CBCT images from a prototype C-arm to simulate surgical deformation and excision.
Main Results:
- The modified algorithm successfully accounted for tissue excision during the registration process.
- Registration performance was quantified, demonstrating accurate handling of anatomical changes.
- The method avoided spurious deformations around excised tissue areas, ensuring registration integrity.
Conclusions:
- The proposed deformable registration method effectively handles excised tissue in intraoperative imaging.
- This advancement is well-suited for improving CBCT-guided head and neck surgical procedures.
- The technique offers enhanced accuracy and reliability for real-time surgical navigation.
Keywords:
3D imagingC-armcone-beam CTflat-panel detectorhead and neck surgeryimage registrationimage-guided interventionssurgical navigation