Evaluation of Intraoperative Volumetric Assessment of Breast Volume Using 3D Handheld Stereo Photogrammetric Device
Rafael Loucas1, Marios Loucas1,2, Sebastian Leitsch1
1Thurgau Hospital Group, Department of Hand and Plastic Surgery, CH-8500 Frauenfeld, Switzerland.
Abstract:
Methods for assessing three-dimensional (3D) breast volume are becoming increasingly popular in breast surgery. However, the precision of intraoperative volumetric assessment is still unclear. Until now, only non-validated scanning systems have been used for intraoperative volumetric analyses. This study aimed to assess the feasibility, handling, and accuracy of a commercially available, validated, and portable device for intraoperative 3D volumetric evaluation. All patients who underwent breast surgery from 2020 to 2022 were identified from our institutional database. Intraoperative 3D volumetric assessments of 103 patients were included in this study. Standardized 3D volumetric measurements were obtained 3 months postoperatively to compare the intraoperatively generated volumetric assessment. All of the study participants were women with a mean age of 48.3 ± 14.7 years (range: 20-89). The mean time for intraoperative volumetric assessment was 8.7 ± 2.6 min. The postoperative 3D volumetric assessment, with a mean volume of 507.11 ± 206.29 cc, showed no significant difference from the intraoperative volumetric measurements of 504.24 ± 276.61 cc (p = 0.68). The mean absolute volume difference between the intraoperative simulations and postoperative results was 27.1 cc. Intraoperative 3D volumetric assessment using the VECTRA H2 imaging system seems to be a feasible, reliable, and accurate method for measuring breast volume. Based on this finding, we plan to investigate whether volumetric objective evaluations will help to improve breast symmetry in the future.