Evaluation of the Performance of OTOPLAN-Based Cochlear Implant Electrode Array Selection: A Retrospective Study
Dayse Távora-Vieira1,2,3, Marcus Voola1,2, Jafri Kuthubutheen1,2,4
1Fiona Stanley Fremantle Hospitals Group, Perth, WA 6150, Australia.
Journal of Personalized Medicine
|August 25, 2023
Summary
Otoplan software recommendations for cochlear implant electrode arrays were compared to actual surgical choices. While clinicians favored shorter arrays, longer arrays may improve hearing outcomes with improved pre-operative planning.
Area of Science:
- Otolaryngology
- Medical Imaging
- Surgical Planning
Background:
- Cochlear implant surgery requires precise electrode array selection.
- Otoplan software provides electrode array recommendations based on imaging data.
- Evaluating the impact of software recommendations on surgical choices and patient outcomes is crucial.
Purpose of the Study:
- To compare Otoplan software's electrode array recommendations with actual implanted arrays.
- To assess the influence of Otoplan-recommended versus non-recommended array selection on auditory outcomes.
- To investigate potential biases in electrode array selection during cochlear implant surgery.
Main Methods:
- Retrospective analysis of 114 patients undergoing cochlear implant surgery.
- Pre-operative CT images analyzed using Otoplan software for array recommendations.
- Comparison of recommended array type and length with surgically implanted arrays.
- Evaluation of auditory outcomes (CNC scores) at 12 months post-implantation.
Main Results:
- 47% of patients received Otoplan-recommended arrays; 34% received shorter, 18% received longer arrays.
- 83% received more flexible arrays, often due to cochlear structure or hearing preservation needs.
- A non-significant negative correlation was found between array selection difference and 12-month CNC scores.
Conclusions:
- Clinicians may exhibit a bias towards shorter electrode arrays, potentially impacting optimal hearing results.
- Pre-operative 3D imaging and surgical planning software can enhance confidence in selecting longer arrays when appropriate.
- Utilizing Otoplan recommendations and advanced imaging may lead to improved auditory outcomes in cochlear implant patients.
Keywords:
OTOPLANanatomy-based fittingcochlear implantelectrode arrayhearing outcomespre-operative planning