Adjustable Stiffness, External Fixator for the Rat Femur Osteotomy and Segmental Bone Defect Models
Published on: October 9, 2014
The Influence of Sagittal Pin Angulation on the Stiffness and Pull-Out Strength of a Monolateral Fixator Construct
Axel Klemeit1, Anna Weber2, Christoph Bourauel2
1Department of Orthopedics and Trauma Surgery, University Hospital Bonn, 53127 Bonn, Germany.
Abstract:
Monolateral pin-to-bar-clamp fixators are commonly used to stabilize acute extremity injuries. Certain rules regarding frame geometry have been established that affect construct stability. The influence of sagittal pin angulation on construct stiffness and strength has not been investigated. The purpose of this biomechanical study was to demonstrate the effect of a pin angulation in the monolateral fixator using a composite cylinder model. Three groups of composite cylinder models with a fracture gap were loaded with different mounting variants of monolateral pin-to-bar-clamp fixators. In the first group, the pins were set parallel to each other and perpendicular to the specimen. In the second group, both pins were set convergent each in an angle of 15° to the specimen. In the third group, the pins were set each 15° divergent. The strength of the constructions was tested using a mechanical testing machine. This was followed by a cyclic loading test to produce pin loosening. A pull-out test was then performed to evaluate the strength of each construct at the pin-bone interface. Initial stiffness analyses showed that the converging configuration was the stiffest, while the diverging configuration was the least stiff. The parallel mounting showed an intermediate stiffness. There was a significantly higher resistance to pull-out force in the diverging pin configuration compared to the converging pin configuration. There was no significant difference in the pull-out strength of the parallel pins compared to the angled pin pairs. Convergent mounting of pin pairs increases the stiffness of a monolateral fixator, whereas a divergent mounting weakens it. Regarding the strength of the pin-bone interface, the divergent pin configuration appears to provide greater resistance to pull-out force than the convergent one. The results of this pilot study should be important for the doctrine of fixator mounting as well as for fixator component design.
More Related Videos
06:41A Minimally Invasive Model to Analyze Endochondral Fracture Healing in Mice Under Standardized Biomechanical Conditions
Published on: March 22, 2018
04:19Minimally Invasive Treatment for Thoracolumbar Burst Fracture Using Sagittal Alignment Screws and A Trauma Reduction Device
Published on: November 8, 2024
Related Concept Videos
Relation between Poisson's ratio, Modulus of Elasticity and Modulus of Rigidity
General Case of Eccentric Axial Loading
Consider a member subjected to equal and opposite forces that are applied along a line that does not coincide with the member's neutral axis. In unsymmetrical...
Stability of structures
Normal Strain under Axial Loading
Eccentric Axial Loading in a Plane of Symmetry
Torsion of Noncircular Members