Variability of quantitative automated perimetry in normal observers
Ophthalmology
|July 1, 1986
Summary
Automated static perimeters show test/retest variability in visual field testing. Variability increases with eccentricity, impacting diagnostic accuracy for glaucoma and other conditions.
Area of Science:
- Ophthalmology
- Visual Science
- Medical Device Technology
Background:
- Automated static perimeters are crucial for diagnosing and monitoring visual field defects.
- Understanding test/retest variability is essential for reliable visual field assessment.
Purpose of the Study:
- To evaluate the test/retest variability of six commercial automated threshold static perimeters.
- To compare variability across different regions of the visual field.
Main Methods:
- Six commercially available automated static perimeters were tested.
- Three projection perimeters (Squid, Octopus 500, Humphrey Field analyzer) and three LED perimeters (Dicon 2000, Fieldmaster 50, Digilab 350) were used.
- Normal volunteers participated in the test/retest evaluations.
Main Results:
- All tested perimeters exhibited some degree of test/retest variability.
- Variability was found to increase from the central 10 degrees to 30 degrees of eccentricity for all devices.
- Some devices demonstrated better performance (lower variability) than others.
Conclusions:
- Despite advancements, all automated static perimeters have inherent test/retest variability.
- Increased variability in the mid-peripheral visual field (10-30 degrees) may affect diagnostic sensitivity.
- Clinicians should consider device-specific variability when interpreting visual field test results.