Related Experiment Video
Updated: Jul 18, 2025

Qualitative Identification of Carboxylic Acids, Boronic Acids, and Amines Using Cruciform Fluorophores
Published on: August 19, 2013
Determination of o-phthalaldehyde for dose verification of the clinical disinfectant by fluorescent sequential
Hung-Yu Chen1, Richie L C Chen1, Bo-Chuan Hsieh1
1Department of Biomechatronics Engineering, College of Bio-Resources and Agriculture, National Taiwan University, No. 1, Sec. 4, Roosevelt Road, Taipei, 106, Taiwan.
Abstract:
A new automated, generic analytical approach for determining the clinical disinfectant o-phthalaldehyde (OPA) is reported in this study. The proposed sequential injection analysis (SIA) is based on the online reaction of the OPA with glycine/N-acetylcysteine (NAC) in a neutral medium (pH = 7.0) to form a highly fluorescent isoindole derivative. All critical flow and reaction variables were investigated, while validation was carried out in the linearity detection range (0.0075-0.02%). As a result, excellent linearity (R2 > 0.99) and precision (1.5-2.4% for repeatability and 0.7-2.2% for reproducibility) were achieved for the reference OPA solutions. Furthermore, reasonable concentration verification of OPA disinfection (0.2-0.6%) in healthcare institutes can be achieved using the developed fluorescent SIA due to its good sensitivity (0.111 V/%) and precision (1.0-2.3% for intermediate precision) around the minimum effective concentration (MEC) of 0.3% for Cidex-OPA disinfectant.

