Absorbed dose evaluation of a blood irradiator with alanine, TLD-100 and ionization chamber
S Grasso1, A Varallo2, R Ricciardi2
1Post Graduate School in Medical Physics, University of Naples Federico II, 80131, Naples, Italy.
Abstract:
Irradiation of blood bags using X-ray irradiators and dosimetry services are required to ensure uniform dose levels in the range 25-50 Gy to prevent Transfusion Associated Graft versus Host Disease (TA-GvHD). An absorbed dose characterization of a Raycell MK2 X-Irradiator was performed using three different dosimetric systems. Results showed a dosimetric accuracy of the ionization chamber together with the Alanine dosimeter. TLDs measurements exhibited a small overestimation by 4% of the absorbed dose. The Dose Uniformity Ratio (DUR), between maximum and minimum dose levels in the canister, was in good agreement with the manufacturer specifications (≤1.5).