Introduction:
Dynamization of proximal femoral nailing by removal of distal interlocking is one of the recommended treatment options for nonunions of femur fractures. However, in certain inter-/subtrochanteric fractures, gliding of the nail along the femoral shaft is blocked by lateral femoral cortical support of the lag screw. For these cases, Biber et al. proposed lateral cortical notching (LCN), in which the supporting lateral bone is removed. This study investigates the biomechanical effect of LCN on gliding of proximal femoral nailing and stress distribution at the bone/implant interface.
Materials And Methods:
In this finite element analysis a three-dimensional model of an unstable intertrochanteric fracture with proximal femoral nailing without distal interlocking was simulated using the FebioStudio software suite. To simulate LCN, the lag screw hole was lengthened to 15.34 mm at the lateral cortex. Displacement of the nail along the femoral shaft axis and von Mises stress distribution were compared between LCN model and standard implantation model.
Results:
Displacement of the nail along the femoral shaft axis was higher in the LCN model than in the standard implantation model (0.48 mm vs. 0.07 mm). Highest von Mises stresses of 176-178 MPa at the implant and of 52-81 MPa at the proximal femur were detected. Maximum von Mises stresses of the implant were comparable at all sides, except for a reduced von Mises stress at the lateral inferior side in the LCN model (80 vs. 102 MPa). At the inferior lateral screw hole and the anterior/posterior lateral screw hole maximum von Mises stress was reduced in the LCN model (2 vs. 49 MPa and 52 vs. 81 MPa), whereas the maximum von Mises stress at the inferior medial screw hole was higher in the LCN model than in the standard implantation model (53 vs. 27 MPa).
Conclusions:
Lateral cortical notching facilitates gliding of a distally dynamized proximal femoral nail along the femoral shaft axis in intertrochanteric fractures. Additionally, the lack of lateral cortical bone support at the lag screw reduces von Mises stress at the bone/implant interface and thus could lower the risk for implant breakage and peri‑implant fractures.
More Related Videos
08:04Proximal Cadaveric Femur Preparation for Fracture Strength Testing and Quantitative CT-based Finite Element Analysis
Published on: March 11, 2017
07:41An Intramedullary Locking Nail for Standardized Fixation of Femur Osteotomies to Analyze Normal and Defective Bone Healing in Mice
Published on: November 13, 2016
Related Concept Videos
Normal Strain under Axial Loading
Bones of the Lower Limb: Femur and Patella