A Detailed Protocol for Perspiration Monitoring Using a Novel, Small, Wireless Device
Published on: November 24, 2016
Cloth-based microfluidic devices integrated onto the patch as wearable colorimetric sensors for simultaneous sweat
Bambang Kuswandi1, Lukman H Irsyad1, Ayik R Puspaningtyas1
1Chemo and Biosensors Group, Faculty of Pharmacy, University of Jember, Jl. Kalimantan 37, Jember, East Java, 68121, Indonesia.
A new wearable pain relief patch doubles as a point-of-care device for simultaneous sweat analysis of lactic acid, sodium ions, and pH. This flexible sensor provides accurate, real-time monitoring during daily activities and exercise.
Area of Science:
- Biomedical Engineering
- Analytical Chemistry
- Wearable Technology
Background:
- Point-of-care (POC) devices are crucial for real-time health monitoring.
- Wearable sensors offer a non-invasive approach to analyze physiological biomarkers.
- Integrating multiple sensing capabilities into a single wearable device presents a significant advancement.
Purpose of the Study:
- To develop a flexible, wearable point-of-care (POC) device integrated into a pain relief patch.
- To enable simultaneous colorimetric sweat analysis for lactic acid, sodium ions, and pH.
- To validate the device's functionality during exercise and daily activities.
Main Methods:
- Development of cloth-based microfluidic devices (CMDs) using wax printing technology on cotton cloth.
- Immobilization of sensitive reagents in sensing zones for micro-volume sweat sample analysis.
- Quantitative and semi-quantitative colorimetric analysis using a smartphone camera and a free app (Color Grab).
Main Results:
- Excellent linear correlation between ∆RGB values and analyte concentrations for lactic acid (R² = 0.994), sodium ion (R² = 0.998), and pH (R² = 0.994).
- Achieved limits of detection (LOD) of 45.73 µg/mL for lactic acid and 56.46 µg/mL for sodium ions.
- Demonstrated high reproducibility for lactic acid (0.79%) and sodium ions (0.89%).
Conclusions:
- The developed wearable colorimetric sensors on a pain relief patch are effective for simultaneous sweat analysis.
- The device shows agreement with standard clinical laboratory methods for sweat analysis during exercise.
- This integrated patch offers a promising solution for continuous, non-invasive monitoring of key sweat analytes.