High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
Matrix Effects of Urine Marker Substances in LC-MS/MS Analysis of Drug of Abuse
Bernd Huppertz1, Silke Möller-Friedrich1, Klaus Baum2
1MVZ Labor Dr. Quade & Kollegen, Cologne, Germany; and.
Background:
Analysis of drug abuse is frequently performed using high-performance liquid chromatography with an MS/MS detector and electrospray ionization. In this context, matrix effects, like signal reduction by ion suppression of individual analytes, play an important role. In this study, the authors evaluated the matrix effect caused by polyethylene glycol (PEG) with chain lengths ranging from 6 to 12 repeating units in drug analysis by LC-MS/MS. Selected chain lengths were used in the Ruma urine marker system.
Methods And Results:
Amphetamines, opiates, opioids, antidepressants, psychotics, benzodiazepines, z-substances, and individual drugs, including THCCOOH, cocaine, LSD, and some of their metabolites were investigated. The matrix effect was investigated at PEG concentrations of 500 mcg/mL and 20 mcg/mL. The effect of each PEG molecule was determined. Furthermore, the effects of different common sample preparations on the PEG matrix effects were evaluated. There was a strong correlation between the retention time of PEG and the drug that was ion-suppressed by PEG. The matrix effect decreased to the point where it was within an acceptable range at the lower PEG concentrations investigated in this study.
Conclusions:
Matrix effects were observed for drugs with approximately the same retention times as the individual PEGs. The influence of the different workup methods was not as clear, which may be because of the similar solubilities of the PEGs and some analytes. At low PEG concentrations, the matrix effect was always below 60%, except for nortilidine. All the drugs were detectable. The effect on quantification was less than 15% for substances with deuterated analytes as internal standards and less than 32% for analytes without their own internal standards.
More Related Videos
10:13Author Spotlight: An Efficient Methodology to Confidently Differentiate and Characterize Fentanyl Analogs
Published on: November 8, 2024
08:41Integration of Miniaturized Solid Phase Extraction and LC-MS/MS Detection of 3-Nitrotyrosine in Human Urine for Clinical Applications
Published on: July 14, 2017
Related Concept Videos
Urine Studies I: Urinalysis
High-Performance Liquid Chromatography: Types of Detectors
Inductively Coupled Plasma-Mass Spectrometry (ICP-MS): Interferences
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...