Bone-anchored annular closure device leading to histiocytic-inflammation-induced neuropathy with resolution after
Jesse J McClure1, Mark E Jentoft2, Sukhwinder S Sandhu3
1Department of Neurosurgery, University of Virginia, 200 Jeanette Lancaster Way, Charlottesville, VA, 22901, USA. jm4wc@virginia.edu.
Purpose:
To determine and report the underlying cause of local inflammation causing recurrent neuropathy and multiple operations in a patient with a Barricaid® device.
Methods:
After removal of this patient's Barricaid® device, we sent local inflammatory tissue to pathology for histochemical analysis. Upon discovery of giant cells formation with polarizable foreign bodies, we performed a literature review regarding the Barricaid® device and its elements.
Results:
After two previous operations and three trials of conservative management, the presented patient underwent an L5/S1 TLIF with removal of her previously installed Barricaid® device. There were no signs of device instability/failure nor were there obvious signs of infection. Inflamed tissue proximal to the Barricaid® device was discovered, debrided, and sample sent to pathology. Removal of the Barricaid® device led to subsequent and durable relief of her symptoms. During review of this case, we discovered the polyethylene terephthalate (PET) weave used in the Barricaid® device is known to induce foreign body reactions, and this precise finding was seen in the majority of animal data submitted to the FDA for the device's acceptance.
Conclusion:
Given the constellation of this patient's symptoms, imaging, intraoperative, and pathology findings, previously published reports, and pre-approval data submitted to the FDA, we conclude that the inflammatory response to the PET weave in this patient's Barricaid® device was the ultimate cause of her continued neuropathy despite multiple prior surgical interventions.